Skip to main content

In vitro measurements of physiological glucose concentrations in biological fluids using mid-infrared spectroscopy

Author(s): Liakat, S; Bors, KA; Huang, T-Y; Michel, APM; Zanghi, E; et al

Download
To refer to this page use: http://arks.princeton.edu/ark:/88435/pr1th65
Abstract: Mid-infrared transmission spectroscopy using broadband midinfrared or Quantum Cascade laser sources is used to predict glucose concentrations of aqueous and serum solutions containing physiologically relevant amounts of glucose (50-400 mg/dL). We employ partial least squares regression to generate a calibration model using a subset of the spectra taken and to predict concentrations from new spectra. Clinically accurate measurements with respect to a Clarke error grid were made for concentrations as low as 30 mg/dL, regardless of background solvent. These results are an important and encouraging step in the work towards developing a noninvasive in vivo glucose sensor in the mid-infrared.
Publication Date: 11-Jun-2013
Electronic Publication Date: 11-Jun-2013
Citation: Liakat, S, Bors, KA, Huang, T-Y, Michel, APM, Zanghi, E, Gmachl, CF. (2013). In vitro measurements of physiological glucose concentrations in biological fluids using mid-infrared spectroscopy
DOI: 10.1364/BOE.4.001083
Type of Material: Journal Article
Journal/Proceeding Title: Biomedical Optics Express
Version: Final published version. Article is made available in OAR by the publisher's permission or policy.