Skip to main content

Simple Experimental Methods for Determining the Apparent Focal Shift in a Microscope System

Author(s): Bratton, Benjamin P; Shaevitz, Joshua W.

Download
To refer to this page use: http://arks.princeton.edu/ark:/88435/pr1mg74
Abstract: Three-dimensional optical microscopy is often complicated by a refractive index mismatch between the sample and objective lens. This mismatch causes focal shift, a difference between sample motion and focal-plane motion, that hinders the accuracy of 3D econstructions. We present two methods for measuring focal shift using fluorescent beads of different sizes and ring-stained fluorescent beads. These simple methods are applicable to most situations, including total internal reflection objectives and samples very close to the interface. For distances 0–1.5 μm into an aqueous environment, our 1.49-NA objective has a relative focal shift of 0.57 ± 0.02, significantly smaller than the simple n2/n1 approximation of 0.88. We also expand on a previous sub-critical angle theory by means of a simple polynomial extrapolation. We test the validity of this extrapolation by measuring the apparent focal shift in samples where the refractive index is between 1.33 and 1.45 and with objectives with numerical apertures between 1.25 and 1.49.
Publication Date: 13-Aug-2015
Electronic Publication Date: 13-Aug-2015
Citation: Bratton, Benjamin P, Shaevitz, Joshua W. (2015). Simple Experimental Methods for Determining the Apparent Focal Shift in a Microscope System. PLOS ONE, 10 (8), e0134616 - e0134616. doi:10.1371/journal.pone.0134616
DOI: doi:10.1371/journal.pone.0134616
EISSN: 1932-6203
Pages: e0134616 - e0134616
Type of Material: Journal Article
Journal/Proceeding Title: PLOS ONE
Version: Final published version. This is an open access article.



Items in OAR@Princeton are protected by copyright, with all rights reserved, unless otherwise indicated.